Analizzatore Ossigeno / Azoto / Idrogeno ELEMENTRAC ONH-p Dati tecnici
| Measured elements | oxygen, nitrogen, hydrogen |
|---|---|
| Samples | inorganic |
| Furnace alignment | vertical |
| Sample carrier | graphite crucibles |
| Field of application | ceramics, engineering / electronics, steel / metallurgy |
| Furnace | electrode impulse furnace (max. 8,5 KW*), temperatures in excess of 3,000 °C |
| Detection method | solid state infrared absorption for oxygen thermal conductivity for nitrogen and hydrogen |
| Typical analysis time | 120 - 180 s |
| Chemicals required | copper oxide, magnesium perchlorate, Schuetze reagents,sodium hydroxide |
| Gas required | compressed air, helium 99.995 % pure, nitrogen 99.995% pure, argon 99.995% pure (if required), all gases with (2 - 4 bar / 30 - 60 psi) |
| Power requirements | 3~ 400 V, 50/60 Hz, max. 8,500 W |
| Dimensions (W x H x D) | 56 x 78 x 64 cm |
| Weight | ~ 165 kg |
| Required equipment | PC, monitor, balance (resolution 0.0001g) |
| Optional accessories | carrier gas purification, gas calibration unit, external chiller |
| - | limited to 6,8 kw in application settings |