酸素/水素分析計 ELEMENTRAC OH-p 製品仕様


Measured elements oxygen, hydrogen
Samples inorganic
Furnace alignment vertical
Sample carrier graphite crucibles
Field of application ceramics, engineering / electronics, steel / metallurgy
Furnace electrode impulse furnace (max. 8,5 KW*), temperatures in excess of 3,000 °C
Detection method solid state infrared absorption for oxygen, thermal conductivity for hydrogen
Typical analysis time 120 - 180 s
Chemicals required magnesium perchlorate, Schuetze reagents,sodium hydroxide
Gas required compressed air, nitrogen 99.995 % pure, all gases with (2 - 4 bar / 30 - 60 psi)
Power requirements 3~ 400 V, 50/60 Hz, max. 8,500 W
Dimensions (W x H x D) 56 x 78 x 64 cm
Weight ~ 165 kg
Required equipment PC, monitor, balance (resolution 0.0001g)
Optional accessories carrier gas purification, gas calibration unit, external chiller
- * limited to 6.8 kw in application settings